[Finite element study on calcium phosphate ceramic screw implanting after removing dynamic hip screw]

[拔除动力髋螺钉后植入磷酸钙陶瓷螺钉的有限元研究]

阅读:1

Abstract

OBJECTIVE: To investigate the validity of improving the femur's mechanical characteristics by implanting calcium phosphate ceramic screws after removing dynamic hip screw (DHS). METHODS: The three dimensional finite element model of the femur was built based on the CT scanning of a normal male volunteer. Then the models of the femur with and without DHS were established. According to calcium phosphate ceramic screws with porosity and apparent elastic modulus, 80% and 0.1 GPa were set as group A, 50% and 1.0 GPa as group B, and 30% and 1.5 GPa as group C. Von Mises stress distribution and maximum stress were recorded when the joint was maximally loaded in a gait cycle. RESULTS: The Von Mises in normal femoral shaft was uniform; no phenomena of stress concentration was observed and the maximum stress located at the joint load-bearing site of the proximal femur. The stress concentration was observed in the femur without DHS, and the maximum stress located at the distal femur around the screw hole. By comparing several different calcium phosphate ceramic screws, the stress distribution of group B was similar to normal femur model, and the maximum stress located at the joint load-bearing site. The other screws of groups A and C showed varying degrees of stress concentration. CONCLUSION: Implanting calcium phosphate ceramic screw can improve the mechanical characteristics of the femur after removing dynamic hip screw, and the calcium phosphate ceramic screw with 50% porosity and 1.0 GPa apparent elastic modulus is suitable for implanting.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。