Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition

双能CT定量成像元素组成和质量分数:三材料分解

阅读:9
作者:Xin Liu, Lifeng Yu, Andrew N Primak, Cynthia H McCollough

Abstract

In principle, dual-energy CT can only accurately decompose a mixture into two materials. To decompose a mixture into three constitute materials using dual-energy CT measurements, a third criteria must be provided to solve for three unknowns with only two spectral measurements. One solution is to assume that the sum of the volumes of three constituent materials is equivalent to the volume of the mixture (i.e., volume conservation), but this is not always true. A more generalized solution is to use the principle of mass conservation, which assumes that the sum of the masses of the three constituent materials is equivalent to the mass of the mixture. In this article, a mass-conservation based, three-material decomposition dual-energy CT algorithm is described and experimental validation of the accuracy of the technique presented. The results demonstrate that the proposed method can accurately measure elemental concentrations under low noise imaging conditions. Clinically, this may be applied to measure the mass fraction of any chemical element in a three-material mixture of solutions without the requirement of volume conservation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。