A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development

一例广泛多孔涂层、远端固定式无骨水泥长柄假体植入物断裂的病例报告:假体柄弯曲发展的详细过程

阅读:1

Abstract

We report the fracture of a cementless long stem in a distally fixated, extensively porous-coated femoral prosthesis used for revisional total hip arthroplasty (THA) 9 years previously in a 48-year-old woman, measuring 58 kg in weight and 155 cm in height. Following resolution of an infection 7 years after the revisional THA, a series of posterior dislocations occurred up until 7 months before sudden stem fracture. Extensive radiographic imaging evidence indicated a gradual progression of bending in the stem, and scanning electron microscope energy dispersive X-ray spectroscopy revealed oxygen in the medial and lateral sections, suspecting in vivo corrosion. We retrospectively examined the detailed course of stem bending development prior to stem fracture. The stem bending immediately after the revisional THA, at the first dislocation, and immediately before the stem fracture was 0°, 1.9°, and 5.2°, respectively. We consider that the cyclic loading with poor proximal bone support, distal fixation, and small diameter were the potential risk of this implant fracture. However, the course of stem bending development suggested that repeated operations and several dislocations might have aggravated the implant fatigue in the present case.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。