The rapid detection of procalcitonin in septic serum using immunoaffinity MALDI chips

免疫亲和MALDI芯片快速检测脓毒症患者血清降钙素原

阅读:5
作者:Josef Dvorak, Jana Novakova, Lucie Kraftova, Vendula Studentova, Martin Matejovic, Jaroslav Radej, Thomas Karvunidis, Jan Horak, Marcela Kralovcova, Jaroslav Hrabak, Zuzana Kalaninova, Michael Volny, Petr Novak, Petr Pompach

Background

Sepsis is a common worldwide health condition with high mortality. It is caused by a dysregulated immune response to the pathogen. Severe infections resulting in sepsis can be also determined by monitoring several bloodstream biomarkers, one of them being pro-hormone procalcitonin (PCT). PCT concentration in the bloodstream correlates well with sepsis and in severe cases increases up to a thousand times from the healthy physiological values in a short time. In this study, we developed a rapid technique for PCT detection by MALDI-TOF mass spectrometry, that uses in-situ enrichment directly on the specialized immuno MALDI chips that are utilized as MALDI plates. The method's ability to detect PCT was confirmed by comparing the

Conclusions

Procalcitonin from human serum was successfully enriched and detected using immunoaffinity MALDI chips. The intact PCT was characterized in 13 septic patients. The method is more specific compared to non-MS-based immunoaffinity techniques and allows observation of different variants of PCT in septic patients.

Methods

The MALDI chips used for the detection of PCT were prepared by ambient ion soft landing of anti-PCT antibody on an ITO glass slide. The chips were used for the development of the rapid MALDI-TOF MS method. A parallel method based on affinity enrichment on magnetic beads followed by LC-MS/MS data-dependent peptide microsequencing was used to prove PCT presence in the sample. All samples were also tested by ELISA to determine PCT concentration prior to analyzing them by mass spectrometry methods.

Results

The MALDI chip method was optimized using recombinant PCT spiked into the human serum. The PCT detection limit was 10 ng/mL. The optimized method was used to analyze 13 sera from patients suffering sepsis. The PCT results were confirmed by LC-MS/MS. The measurement of the intact PCT by the MALDI chip method revealed that sera of patients with severe sepsis have other forms of PCT present, which show post-processing of the primary sequence by cleavage of PCT, resulting in the formation of N and C termini fragments. Conclusions: Procalcitonin from human serum was successfully enriched and detected using immunoaffinity MALDI chips. The intact PCT was characterized in 13 septic patients. The method is more specific compared to non-MS-based immunoaffinity techniques and allows observation of different variants of PCT in septic patients.