Bone graft incorporation failure with inappropriate limb load transfer can lead to aseptic acetabular loosening of metal-on-metal prosthesis: A case report

骨移植失败伴随不当的肢体负荷转移可导致金属对金属假体髋臼无菌性松动:病例报告

阅读:1

Abstract

BACKGROUND: Aseptic acetabular loosening can result from various factors that can be categorized into groups: patient-related, surgeon-related and implant-related. We present a case of a 63-year-old patient who at first underwent a total hip arthroplasty (THA) using a metal-on-metal bearing due to hip arthrosis. Follow-up visits revealed no complications after the procedure. Two years after the THA, acetabular component loosening occurred due to subsequent trauma of the opposite hip, necessitating a revision THA using a ceramic-on-ceramic bearing. CASE SUMMARY: We aim to illustrate a rare case where the primary reason for undergoing THA revision was not only incomplete bone graft incorporation but also improper limb load distribution. Following the revision arthroplasty, a 9-year follow-up visit revealed improvements in all evaluation measures on questionnaire compared to the state before surgery: Harris Hip Score (before surgery: 15; after surgery: 95), Western Ontario and McMaster Universities Arthritis Index (before surgery: 96; after surgery: 0), and Visual Analogue Scale (before surgery: 10; after surgery: 1). CONCLUSION: Opposite-hip trauma caused a weight transfer to the limb after a THA procedure. This process led to a stress shielding effect, resulting in acetabular component loosening.