Enhancing sensitivity in atomic force microscopy for planar tip-on-chip probes

提高平面芯片探针原子力显微镜的灵敏度

阅读:1

Abstract

We present a new approach to tuning-fork-based atomic force microscopy for utilizing advanced "tip-on-chip" probes with high sensitivity and broad compatibility. Usually, such chip-like probes with a size reaching 2 × 2 mm(2) drastically perturb the oscillation of the tuning fork, resulting in poor performance in its intrinsic force sensing. Therefore, restoring initial oscillatory characteristics is necessary for regaining high sensitivity. To this end, we developed a new approach consisting of three basic steps: tuning-fork rebalancing, revamping holder-sensor fixation, and electrode reconfiguration. Mass rebalancing allows the tuning fork to recover the frequency and regain high Q-factor values up to 10(4) in air and up to 4 × 10(4) in ultra-high vacuum conditions. The floating-like holder-fixation using soft wires significantly reduces energy dissipation from the mounting elements. Combined with the soft wires, reconfigured electrodes provide electrical access to the chip-like probe without intervening in the force-sensing signal. Finally, our easy-to-implement approach allows converting the atomic force microscopy tip from a passive tool to a dedicated microdevice with extended functionality.