Accuracy of the volume and pressure displays of high frequency oscillators

高频振荡器的体积和压力显示的精度

阅读:1

Abstract

OBJECTIVE: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators. METHODS: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5-15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator's patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz. RESULTS: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators. CONCLUSION: Frequency affects the accuracy of displayed volumes and, to a lesser extent, displayed pressures of high frequency oscillators. The results emphasise that data displayed by new devices should not be uncritically accepted.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。