BACKGROUND: Implants for fracture and/or osteotomy fixation are often tested according to basic mechanical test models such as open gap tests or 4-point-bending tests. These may be suitable to test and compare different implants for safety and clinical approval, but are not always representative of the post-operative situation, which is decisive when it comes to bone healing. In the current study the Knee Expert Group of the Association for the Study of Internal Fixation has compared the available open gap test results of the latest version of the TomoFix Medial Distal Femoral Plate and the antecedent plate design, with the test results of a more physiological and life-like test model. In the open gap test model the antecedent plate design was found to have superior stiffness and fatigue strength. METHODS: In the current study simulated postoperative conditions for medial closing wedge supracondylar osteotomies were used. The constructs were subjected to cyclical axial and torsional loading and were subsequently tested to failure. RESULTS: The more life-like tests in this study showed that the latest version was either more or equally stable and stiff than the antecedent version of the plate, in all of the tests. It is argued that the difference in results between the two loading models is due to differences in test design. CONCLUSIONS: These test results stress the importance of not only using standard open gap and 4-point-bending tests, but also to use as life-like as possible test conditions for any form of biomechanical testing of new implants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。