Electrical Field Interactions during Adjacent Electrode Stimulations: eABR Evaluation in Cochlear Implant Users

相邻电极刺激期间的电场相互作用:人工耳蜗植入者的eABR评估

阅读:1

Abstract

The present study investigates how electrically evoked Auditory Brainstem Responses (eABRs) can be used to measure local channel interactions along cochlear implant (CI) electrode arrays. eABRs were recorded from 16 experienced CI patients in response to electrical pulse trains delivered using three stimulation configurations: (1) single electrode stimulations (E11 or E13); (2) simultaneous stimulation from two electrodes separated by one (E(n) and E(n+2), E11 and E13); and (3) stimulations from three consecutive electrodes (E11, E12, and E13). Stimulation level was kept constant at 70% electrical dynamic range (EDR) on the two flanking electrodes (E11 and E13) and was varied from 0 to 100% EDR on the middle electrode (E12). We hypothesized that increasing the middle electrode stimulation level would cause increasing local electrical interactions, reflected in characteristics of the evoked compound eABR. Results show that group averaged eABR wave III and V latency and amplitude were reduced when stimulation level at the middle electrode was increased, in particular when stimulation level on E12 reached 40, 70, and 100% EDR. Compound eABRs can provide a detailed individual quantification of electrical interactions occurring at specific electrodes along the CI electrode array. This approach allows a fine determination of interactions at the single electrode level potentially informing audiological decisions regarding mapping of CI systems.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。