Evaluating photoneutron dose equivalents for lung cancer using PMMA phantoms undergoing 15 MV IMRT

利用接受 15 MV IMRT 治疗的 PMMA 体模评估肺癌的光中子剂量当量

阅读:1

Abstract

BACKGROUND: According to statistics of the Ministry of Health and Welfare in 2017, the second leading cause of death in Taiwan was lung cancer. OBJECTIVE: Routine treatment planning does not consider photoneutron dose equivalent (PNDE) of patient induced secondary radiation resulting from primary exposure of lung cancer. However, such treatment is potentially important for improving estimates of health risks. METHODS: This study used 10, 30, 50, 70, and 90 kg of polymethylmethacrylate (PMMA) phantoms as patient to measure PNDE varying anatomical area during lung cancer of intensity modulated radiotherapy (IMRT) treatment. Paired thermoluminescent dosimeters (TLD-600 and 700) were calibrated using university reactor neutrons. TLDs were inserted into phantom which was closely corresponded of the represented tissues or organs. RESULTS: Neutron doses (ND) of organ or tissue (N⁢DT) were determined in these phantoms using paired TLDs approach. The risks of incurring fatal secondary malignancies, maximum statistical and total errors were estimated. Evaluated PNDE ranged from 0.80 ± 0.12 to 0.56 ± 0.08 mSv/Gy for these phantoms. CONCLUSION: The estimated N⁢DT decreased with increasing distance that is from the central axis. Evaluated PNDE and N⁢D𝑠𝑘𝑖𝑛 for these phantoms were discussed. This investigation also identified secondary risks associated with PNDE relating to radiation protection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。