Comparison of treatment planning for stereotactic radiosurgery and stereotactic body radiation therapy techniques with 2.5mm and 5mm multileaf collimator (MLC): A pilot study

采用 2.5mm 和 5mm 多叶准直器 (MLC) 进行立体定向放射外科手术和立体定向体部放射治疗技术治疗计划的比较:一项初步研究

阅读:1

Abstract

BACKGROUND: The Elekta micro-multileaf collimator (mMLC) known as apex for planning stereotactic radiosurgery (SRS)/stereotactic radiotherapy (SRT)/stereotactic body radiation therapy (SBRT) provides excellent dose distribution; however, it offers disadvantages such as prolonged treatment duration and technical errors in terms of mMLC and gantry calibration, which adds to the total treatment duration. Hence, we aimed to compare the treatment planning performed with the 2.5mm mMLC and 5mm MLC known as Agility for brain and lung targets treated with SRS and SBRT in Elekta Versa high definition (HD). METHODS: The study included 10 patients, five each with brain and lung targets. Two treatment plans were performed for each case using Elekta's Monaco (5.11.03) treatment planning system (TPS) with 2.5mm and 5mm MLC. An X-ray photon beam of energy 6FFF was used for planning purposes with various gantry, couch, and collimator combinations. These two plans were compared using target coverage (TC), conformity index (CI), homogeneity index (HI), gradient index (GI), and organ at risk (OAR) doses. RESULTS: No significant differences were found in the target coverage, CI, HI, or OAR doses in either MLC design. Volumetric modulated arc therapy (VMAT) with a 5 mm MLC provided equivalent tumor coverage with an additional number of monitor units. OAR doses were comparable in both MLC widths for brain targets, whereas for lung targets, OAR doses were slightly lower with 2.5mm mMLC. GI was superior in the 2.5mm mMLC compared to the 5mm MLC giving a steep falloff in the dose distributions (p = 0.158). CONCLUSIONS: The TC, CI, HI, and OAR doses were similar in both 2.5mm and 5mm based VMAT plans. The gradient index was better in the 2.5mm mMLC resulting in steep dose gradients, which further reduced the isodose volumes. Therefore, a 5mm MLC (agility) can also be used for SRS/SBRT treatment planning, with a further reduction in the gradient index. However, the study must be extended further with more samples and multiple comparison parameters.Clinical Trials Registry - India, registration number CTRI/2021/11/037842, registration date. 8 (th) November, 2021.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。