Sedline(®) Miscalculation of Depth of Anaesthesia Variables in Two Pigs Due to Electrocardiographic Signal Contamination

由于心电信号污染,Sedline®对两头猪麻醉深度变量的计算出现误差

阅读:1

Abstract

Two young (11-week-old) pigs underwent sole propofol anaesthesia as part of an experimental study. The depth of anaesthesia was evaluated both clinically and using the electroencephalography(EEG)-based monitor Sedline; in particular, the patient state index, suppression ratio, raw EEG traces, and its spectrogram were assessed. Physiological parameters and electrocardiographic activity were continuously monitored. In one pig (Case 1), during the administration of high doses of propofol, the Sedline-generated variables suddenly indicated an increased EEG activity while this was not confirmed by observation of either the raw EEG or its spectrogram. In the second pig (Case 2), a similar event was recorded during euthanasia with systemic pentobarbital. Both events happened while the EEG activity was isoelectric except for signal interferences and synchronous in rhythm and shape with the electrocardiographic activity. The suggestion of increased brain activity based on the interpretation of the Sedline variables was suspected wrong; most probably due to electrocardiographic interferences. In pigs, the patient state index and suppression ratio, as calculated by the Sedline monitor, could be influenced by the electrocardiographic activity contaminating the EEG trace, especially during otherwise isoelectric periods (strong EEG depression). Visual interpretation of the raw EEG and of the spectrogram remains necessary to identify such artefacts.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。