Electroplating Deposition of Bismuth Absorbers for X-ray Superconducting Transition Edge Sensors

用于X射线超导转变边缘传感器的铋吸收体的电镀沉积

阅读:1

Abstract

An absorber with a high absorbing efficiency is crucial for X-ray transition edge sensors (TESs) to realize high quantum efficiency and the best energy resolution. Semimetal Bismuth (Bi) has shown greater superiority than gold (Au) as the absorber due to the low specific heat capacity, which is two orders of magnitude smaller. The electroplating process of Bi films is investigated. The Bi grains show a polycrystalline rhombohedral structure, and the X-ray diffraction (XRD) patterns show a typical crystal orientation of (012). The average grain size becomes larger as the electroplating current density and the thickness increase, and the orientation of Bi grains changes as the temperature increases. The residual resistance ratio (RRR) (R(300 K)/R(4.2 K)) is 1.37 for the Bi film (862 nm) deposited with 9 mA/cm(2) at 40 °C for 2 min. The absorptivity of the 5 μm thick Bi films is 40.3% and 30.7% for 10 keV and 15.6 keV X-ray radiation respectively, which shows that Bi films are a good candidate as the absorber of X-ray TESs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。