Comparison between spectral-domain optical coherence tomography-based and partial coherence interferometry-based biometry

基于光谱域光学相干断层扫描和基于部分相干干涉测量的生物测量学比较

阅读:1

Abstract

We compared axial length (AL) measurements using spectral-domain optical coherence tomography (SD-OCT)-based and partial coherence interferometry (PCI)-based biometry in eyes with various chorioretinal diseases. Medical records were retrospectively analyzed in 123 eyes from 73 patients who underwent AL measurements using SD-OCT and PCI. AL values were highly consistent by both measurements (ICC = 0.996 in SD-OCT and 0.995 in PCI-based biometry). The mean difference in AL between the two instruments was - 0.09 ± 0.17 mm, which was significant (P < 0.001). There was a strong correlation between the measurements obtained using the two instruments (r = 0.982, P < 0.001). The 95% limits of agreement were - 0.43 to 0.25 mm, and the agreement of AL in the Bland-Altman analysis showed a difference. The difference in measurements was not associated with the difference in central retinal thickness (r = 0.033, P = 0.720). However, the presence of an epiretinal membrane affected the difference in measurements, unlike in other chorioretinal diseases (P = 0.014). SD-OCT-based biometry showed a good correlation with PCI-based biometry in the measurement of AL; however, the mean AL between the two instruments was significantly different. Therefore, AL measurements made using SD-OCT-based and PCI-based biometry are not interchangeable.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。