Accuracy of a Smart Diaper System for Nursing Home Residents for Automatically Detecting Voided Volume: Instrument Validation Study

智能尿布系统在养老院居民中自动检测排尿量的准确性:仪器验证研究

阅读:1

Abstract

BACKGROUND: Diapers are commonly used by older patients with urination disorders. A smart diaper system (SDS) may be able to estimate the weight of urine comparably to conventional measurements made by weighing diapers. OBJECTIVE: The aim of the current research is to determine the degree of accuracy of an SDS technology specifically designed for the management of urination routines and the use of incontinence pads in older adults. METHODS: From January to December 2022, 97 older patients with at least 1 chronic disease from 3 nursing homes were included. In this study, the SDS was used for 1 month per patient after obtaining their consent; all patients originally used traditional diapers in the nursing home. The index test measured the change in electrical resistance of the SDS and the reference test measured the change in actual urine weight. When measuring the actual urine weight, the degree of absorption was directly confirmed with the naked eye because the expression value varied according to pressure changes caused by the user's movement or position. The Pearson correlation was used to determine the correlation between the 2 test methods, the intraclass correlation coefficient (ICC) was used to check the degree of agreement between the evaluators, and the Bland-Altman test was used to confirm whether there was a significant difference between the 2 test methods. RESULTS: The average age of the 97 participants was 86.48 (SD 6.26) years, with 10 men and 87 women. There were 73 patients (75%) with hypertension, 86 patients (88%) with dementia, and 86 patients (88%) with 2 or more comorbidities, accounting for the majority. The Pearson correlation coefficient and ICC were 0.971 and 0.985 (P<.001). In the Bland-Altman figure, the difference in the mean between the 2 tests was evenly scattered without showing a specific pattern, indicating that the SDS and actual urine weight were very consistent. The difference between the mean of the 2 tests was -0.045 of the standardized mean difference, and all measurements were located within the 95% CI, so this confirms that the 2 test methods are equivalent. CONCLUSIONS: Our study showed a fairly high correlation coefficient and ICC for all patient groups, which reveals that the 2 tests were very consistent and that the SDS can replace traditional diapers, even in a real clinical setting. This study shows the possibility that heath care professionals could be alerted by the SDS to the need for pad replacement due to incontinence, thus avoiding the development of dermatological complications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。