Digital Subtraction Angiography of Cerebral Arteries: Influence of Cranial Dimensions on X-ray Tube Performance

脑动脉数字减影血管造影:颅骨尺寸对X射线管性能的影响

阅读:1

Abstract

(1) Background. Digital subtraction angiography (DSA) is indispensable for diagnosing cerebral aneurysms due to its superior imaging precision. However, optimizing X-ray parameters is crucial for accurate diagnosis, with X-ray tube settings significantly influencing image quality. Understanding the relationship between skull dimensions and X-ray parameters is pivotal for tailoring imaging protocols to individual patients. (2) Methods. A retrospective analysis of DSA data from a single center was conducted, involving 251 patients. Cephalometric measurements and statistical analyses were performed to assess correlations between skull dimensions and X-ray tube parameters (voltage and current). (3) Results. The study revealed significant correlations between skull dimensions and X-ray tube parameters, highlighting the importance of considering individual anatomical variations. Gender-based differences in X-ray parameters were observed, emphasizing the need for personalized imaging protocols. (4) Conclusions. Personalized approaches to DSA imaging, integrating individual anatomical variations and gender-specific differences, are essential for optimizing diagnostic outcomes. While this study provides valuable insights, further research across multiple centers and diverse imaging equipment is warranted to validate these findings.