Effects of timing parameter changes on the gait of functional electrical stimulation users with drop foot

时序参数变化对垂足功能性电刺激使用者步态的影响

阅读:1

Abstract

INTRODUCTION: Functional electrical stimulation uses clinician-set parameters to modify stimulation. This study aimed to investigate whether timing parameters in the ODFS Pace functional electrical stimulation device have an effect on the gait of the general population of functional electrical stimulation users who have a foot drop. METHODS: Twelve functional electrical stimulation users with foot drop resulting from upper motor neurone disorders were recruited from the functional electrical stimulation Service in Leeds, UK. A crossover trial design was used, comparing adjusted values of rising ramp, delay and extension. Instrumented gait analysis was carried out to measure ankle dorsiflexion during the swing phase of gait, foot clearance from the ground, and speed of ankle plantarflexion at initial contact. The effect of timing parameters on gait kinematics was studied. RESULTS: No statistically significant effects on the measured parts of gait were found for any of the timing parameters. Trends were identified in average mid-swing ground clearance and dorsiflexion associated with the delay and rising ramp timing parameters. CONCLUSIONS: Further work in this area should use larger numbers of participants. Based on these results, the effects of ramping and delay would be of particular interest for further study.