Effect of head tilt on repeatability of optic nerve head parameters using cirrus spectral-domain optical coherence tomography

利用Cirrus光谱域光学相干断层扫描技术研究头部倾斜对视神经乳头参数重复性的影响

阅读:1

Abstract

AIM: To assess the repeatability of measuring optic nerve head (ONH) parameters using the Cirrus optical coherence tomography (OCT), as well as to assess the effect of head tilt on these measurements. METHODS: Thirty healthy participants with no evidence of glaucoma were recruited for the study. Visual acuity, intraocular pressure, standard automated perimetry and ocular examination were performed for each participant. One eye was then randomly selected and scanned undilated with the Cirrus OCT in 3 positions (neutral, 30° right tilt and 30° left tilt). RESULTS: Data collected from 29 eyes were used for analysis. One patient was omitted due to poor scan quality. The repeatability of the ONH parameters was analyzed using analysis of variance, coefficient of variation (COV) and intraclass correlation coefficient (ICC). Analysis of variance showed no statistically significant difference between 3 scans in a single position. There was good agreement between measurements (ICC 0.919-0.996, COV 1.94%-5.48%). Even with the presence of head tilt, repeated scans in the 3 positions showed good agreement as well (ICC 0.888-0.996, COV 2.04%-5.39%). CONCLUSION: Serial measurements of ONH parameters using the Cirrus OCT are found to have good repeatability. The ONH parameters with Cirrus OCT also maintain good repeatability despite head tilt.