The effect of flow and pressure on the intraoxygenator flow path of different contemporary oxygenators: an in vitro trial

不同现代氧合器内部流路中流量和压力的影响:一项体外试验

阅读:1

Abstract

INTRODUCTION: This study analyzed the effect of different flows and pressures on the intraoxygenator flow path in three contemporary oxygenators and its consequences for oxygen transfer efficiency. METHODS: In an experimental setup, intraoxygenator flow path parameters were analyzed at post-oxygenator pressures of 150, 200, and 250 mm Hg and at flows ranging from 2 L/min to the oxygenators' maximum permitted flow, with and without pulsatility. The oxygen gradient and the oxygen transfer per minute and per 100 mL blood were calculated using previously collected clinical data and compared with the flow path parameters. RESULTS: Increasing pressure did not affect the flow path parameters, whereas pulsatile flow led to significantly increased dynamic oxygenator blood volumes. Increased flow resulted in decreased values of the flow path parameters in all oxygenators, indicating increased flow through short pathways in the oxygenator. In parallel, oxygen transfer/100 mL blood decreased in all oxygenators (average 2.5 ± 0.4 to 2.4 ± 0.3 mL/dL, p > 0.001) and the oxygen gradient increased from 229 ± 45 to 287 ± 29 mm Hg, p > 0.001, indicating decreased oxygen transfer efficiency. Oxygen transfer/min increased (101 ± 15 to 143 ± 20 mL/min/m(2), p > 0.001), however, due to the increased flow through the oxygenator. CONCLUSION: Varying trans-membrane oxygenator pressures did not lead to changes in the intraoxygenator flow path, while an increased flow exhibited lower flow path parameters resulting in less efficient use of the gas exchange compartment. The latter was confirmed by a decrease in O(2) transfer efficiency during higher blood flows.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。