Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

技术说明:关于 GAFChromic EBT-XD 胶片和横向响应伪影

阅读:1

Abstract

PURPOSE: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. METHODS: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. RESULTS: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. CONCLUSIONS: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film requires higher dose to reach the same response as EBT3, resulting in lesser impact of the lateral response artifact. For doses >10 Gy, the slopes of the EBT-XD red and green channel dose response curves are greater than the corresponding ones for EBT3. For these two reasons, the authors prefer EBT-XD for doses exceeding about 10 Gy.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。