Dimensional compatibility and limitations of tracheal intubation through supraglottic airway devices: a mannequin-based in vitro study

经声门上气道装置进行气管插管的尺寸兼容性和局限性:一项基于人体模型的体外研究

阅读:3

Abstract

BACKGROUND: Flexible bronchoscope-guided tracheal intubation through supraglottic airway devices (SGAs) is a well-established element of difficult intubation algorithms. Success can be limited by dimensional incompatibilities between tracheal tubes (TTs) and SGAs. METHODS: In this in vitro study, we tested the feasibility of TT passage through SGAs, removal of SGAs over TTs, and the ability to guide the flexible bronchoscope with 13 TT brands (internal diameter, 6.5-8.0 mm) and ten different SGAs (#4 and #5) in an intubation mannequin. RESULTS: We tested 1,040 combinations of SGAs and TTs. Tracheal tube passage failed in 155 (30%) combinations of the five tested first-generation SGAs (117 [46%] with SGA #4, 38 [15%] with SGA #5) and in three (0.6%) combinations of the five tested second-generation SGAs (two [0.8%] with SGA #4 and one [0.4%] with SGA #5). The reason for failed passage of a TT through a first-generation SGA consistently was a too-narrow SGA connector. Removal of the SGA over the TT in the 882 remaining combinations was impossible for all sizes of reinforced TTs, except the Parker Reinforced TT, and was possible for all non-reinforced TTs. Only one combination with SGA #4 and 84 combinations with SGA #5 were not ideal to adequately guide the flexible bronchoscope. CONCLUSION: Clinically relevant combinations of adult-size TTs and SGAs can be incompatible, rendering flexible bronchoscope-guided tracheal intubation through an SGA impossible. Additional limitations exist regarding removal of the SGA and maneuverability of the flexible bronchoscope.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。