Technical pitfalls and proposed modifications of instructions for use for endovascular aortic aneurysm repair using the Gore Excluder conformable device in angulated and short landing zones

使用 Gore Excluder 可变形装置在弯曲和短着陆区进行血管内主动脉瘤修复时的技术缺陷及使用说明的拟议修改

阅读:1

Abstract

We describe a case of an abdominal aortic aneurysm (AAA) and angulated proximal neck treated with a Gore Excluder conformable endoprosthesis and show relevant technical pitfalls in the deployment of the graft main body. An 82-year-old man presented with a 71-mm asymptomatic AAA with an angulated infrarenal proximal neck (75°) and was referred to our unit. The patient was treated with a 26-mm Gore Excluder conformable device, which was deployed slightly above the renal arteries after precatheterization of the lowest renal artery. The graft was then repositioned with support of the introducer sheath and a stiff guide wire. The proximal sealing zone was ballooned before the endograft delivery system was retrieved to avoid distal migration. Technical success was achieved. The patient was discharged with no complications. No type Ia endoleak was present on the 6-month computed tomography scan. Endovascular treatment of an AAA with a severe angulated proximal neck can be effective with a conformable stent graft if technical measures are used during deployment of the main body to optimize the seal.