The impact of off the visual axis retinoscopy on objective central refractive measurement in adult clinical practice: a prospective, randomized clinical study

非视轴视网膜镜检查对成人临床实践中客观中心屈光度测量的影响:一项前瞻性随机临床研究

阅读:1

Abstract

PURPOSE: The objective of this study was to examine the effect of the off axis retinoscopy on objective central refractive measurement in adult clinical practice. METHODS: In all, 40 subjects underwent undilated retinoscopy in a randomly selected eye both on and off the visual axis by a single masked examiner. Off axis retinoscopy was defined as retinoscopy performed with the testing eye of the examiner aligned with the contralateral (non-test) eye of the subject resulting in an off axis deviation in the nasal horizontal visual field. Retinoscopy was performed in negative cylinder only and spherocylindrical measurements were converted to power vectors for analysis. Paired t-test was used to assess differences in M, J(0) and J(45) power vectors including differences between mean aided and unaided LogMar acuities. RESULTS: In all, 14 subjects were myopic (SE≤-0.5 D), 13 subjects were emmetropic (SE between -0.49 and 1.0 D) and 13 subjects were hyperopic (SE >1.0 D). Mean angle of deviation was 5.58° in the nasal horizontal visual field. J(0) showed a significant negative shift in those with myopia (P<0.001) and emmetropia (P=0.049) following off axis retinoscopy. No significant differences in M, J(0) and J(45) were found in the hyperopes. Mean aided LogMar acuities after on and off axis retinoscopy were both significantly better than mean unaided LogMar VA (P<0.001 in both cases). CONCLUSION: Small degrees of off axis retinoscopy encountered in everyday clinical practice can induce errors in objective central refractive measurement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。