Comparison of INTEGRA and the Manual Method to Determine the Axis for Intraocular Lens Implantation-A Case Series of 60 Eyes

INTEGRA 与手动方法在确定人工晶状体植入轴位方面的比较——60 只眼的病例系列研究

阅读:1

Abstract

(1) Background: To compare the results of a new intraoperative contactless device (INTEGRA Optomed, Poland) with the result of a manual method for determining the axis for toric intraocular lens implantation. (2) Material and Methods: This retrospective observational study included 60 eyes of 40 patients (17 men, 23 women) who had toric intraocular lenses implanted. A video recording of each surgery that used the INTEGRA system was made for the analysis. Two researchers then independently assessed the location of the implant axes determined with both digital and manual slit-lamp methods, and compared the results between methods. (3) Results: The implantation axes suggested through the manual and INTEGRA methods were similar. The median axis disparities were 0.0 degrees for both groups. The standard deviation was 0.63 and 0.75 for researcher 1 and 2, respectively. The dominant value was 0.0 in both groups. The INTEGRA axis designation was statistically significantly different from the manual method for researcher 1 (p < 0.05), but it was statistically insignificant for researcher 2 (p = 0.79). (4) Conclusions: The INTEGRA system is a digital ink-free device for image tracking scleral vessels. It was helpful for determining the implantation axis in a precise manner, and the measurements were comparable with those obtained through a manual technique.