Fluid warming with parylene-coated enFlow cartridge: Bench and pilot animal study of aluminum extraction due to prolonged use

使用聚对二甲苯涂层 enFlow 滤芯进行液体加温:长期使用导致铝提取的实验室研究和动物试验

阅读:1

Abstract

OBJECTIVES: Intravenous fluid warming devices with surface heating systems transfer heat using aluminum blocks, which if uncoated elute toxic levels of aluminum into the infusate. This study examined extractable aluminum detected from prolonged use of the updated version of the enFlow(®) cartridge, which uses a parylene-coated aluminum heating block. METHODS: In dynamic bench tests, we measured the concentration of aluminum that leached into three solutions (Sterofundin ISO, Plasma-Lyte 148, and whole blood) that were continuously pumped (0.2 and 5.5 mL min(-1)) and warmed to 40°C by the enFlow cartridge (parylene-coated) for 5 h. Prolonged quasi-static bench tests measured aluminum concentration in 16 solutions which were gently rocked within the enFlow cartridge (parylene-coated) for 72 h at 40°C. Aluminum concentrations were measured using inductively coupled mass spectroscopy and matrix blank corrected. Measured aluminum concentrations were compared to a Tolerable Exposure limit to calculate Margins of Safety based on the US Food and Drug Administration maximum recommended concentration in parenteral fluids (25 μg L(-1)). A parallel pilot in vivo animal study was performed using mice injected with fluids warmed for 72 h by the enFlow cartridge (parylene-coated). RESULTS: The enFlow cartridge (parylene-coated) demonstrated low toxicological risks in all tests. Sterofundin ISO resulted in the highest aluminum concentration after simulated prolonged use of the enFlow cartridge (parylene-coated) (3.11 μg device(-1)), which represents a 99.2% decrease from the enFlow cartridge (uncoated) and Margin of Safety of 1.7. Dynamic tests at two different flow rates with three challenge solutions resulted in concentrations less than the method detection limits (20.6 or 41.2 μg L(-1)) of the analysis method. The animals in the in vivo study showed no evidence of toxicity. CONCLUSION: Observed toxicological risk levels associated with the enFlow cartridge (parylene-coated) intravenous fluid warmer were below those set by the Food and Drug Administration and suggest that the use of enFlow cartridge (parylene-coated) is safe with a variety of intravenous solution types and in different therapeutic scenarios.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。