Comparison of sequentially measured Aloka echo-tracking one-point pulse wave velocity with SphygmoCor carotid-femoral pulse wave velocity

比较 Aloka 回声追踪单点脉搏波速度与 SphygmoCor 颈股动脉脉搏波速度的顺序测量值

阅读:1

Abstract

OBJECTIVES: Recently, echo-tracking-derived measures of arterial stiffness have been introduced in clinical practice for the assessment of one-point pulse wave velocity. The purpose of this study was to find a relation between carotid-femoral pulse wave velocity and one-point carotid pulse wave velocity, and to find a value of one-point carotid pulse wave velocity that predicts carotid-femoral pulse wave velocity higher than 12 m/s. METHODS: A total of 160 consecutive subjects (112 male/48 female, mean age = 51.5 ± 14.1 years; 96 healthy, 44 hypertensives, 13 with aortic valve disease, and 7 with left ventricular dysfunction) were studied. Carotid-femoral pulse wave velocity was measured with the SphygmoCor system and one-point carotid pulse wave velocity with high-definition echo-tracking system (ProSound Alpha10; Aloka, Tokyo, Japan). RESULTS: Both carotid-femoral pulse wave velocity and one-point carotid pulse wave velocity correlated significantly with each other (r = 0.539, p < 0.001) and with age (one-point carotid pulse wave velocity r = 0.618, carotid-femoral pulse wave velocity r = 0.617, p < 0.0001 for both). Median value of carotid-femoral pulse wave velocity (7.2 m/s, 95% confidence interval = 6.2-8.9) was systematically higher than that of one-point carotid pulse wave velocity (5.8 m/s, 95% confidence interval = 5-6.6). The area under the receiver operating characteristic curve was 0.85, identifying the cutoff for one-point pulse wave velocity of 6.65 m/s as the best predictor of carotid-femoral pulse wave velocity more than 12 m/s (sensitivity = 0.818, specificity = 0.819). CONCLUSIONS: One-point carotid pulse wave velocity correlates with carotid-femoral pulse wave velocity, and the cutoff of 6.65 m/s was the best predictor of carotid-femoral pulse wave velocity over 12 m/s.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。