Clinically Available Low Intensity Ultrasound Devices do not Promote Axonal Regeneration After Peripheral Nerve Surgery-A Preclinical Investigation of an FDA-Approved Device

临床上可用的低强度超声设备并不能促进周围神经手术后的轴突再生——一项针对FDA批准设备的临床前研究

阅读:1

Abstract

The slow axonal regeneration and consecutive delayed muscle reinnervation cause persistent functional deficits following peripheral nerve injury, even following sufficient surgical nerve reconstruction. Preclinically, adjunct ultrasound therapy has shown to significantly accelerate nerve regeneration and thereby improve muscle function compared to nerve reconstruction alone. However, although FDA-approved and clinically well-tested ultrasound devices for other conditions such as delayed-healing fractures are available, they have not been investigated for peripheral nerve injury yet. Aiming to provide a fast clinical translation, we evaluated EXOGEN (Bioventus LLC, Durham, USA), a clinical device for low-intensity ultrasound therapy in various treatment intervals following peripheral nerve surgery. Sixty rats, randomized to five groups of twelve animals each, underwent median nerve transection and primary epineural nerve reconstruction. Post-surgically the ultrasound therapy (duration: 2 min, frequency: 1.5 MHz, pulsed SATA-intensity: 30 mW/cm(2), repetition-rate: 1.0 kHz, duty-cycle: 20%) was applied either weekly, 3 times a week or daily. A daily sham-therapy and a control-group served as references. Functional muscle testing, electrodiagnostics and histological analyses were used to evaluate nerve regeneration. The post-surgically absent grip strength recovered in all groups and increased from week four on without any significant differences among groups. The weekly treated animals showed significantly reduced target muscle atrophy compared to sham-treated animals (p = 0.042), however, with no significant differences to three-times-a-week-, daily treated and control animals. The number of myelinated axons distal to the lesion site increased significantly in all groups (p < 0.001) without significant difference among groups (p > 0.05). A full recovery of distal latency was achieved in all groups and muscle function and CMAP recurred with insignificant differences among groups. In conclusion, the clinically available FDA-approved ultrasound device did not promote the axonal regeneration following nerve injury in comparison to control and sham groups. This is in contrast to a conclusive preclinical evidence base and likely due to the insufficient ultrasound-intensity of 30 mW/cm(2). We recommend the clinical investigation of 200-300 mW/cm(2).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。