Continuous Intraoperative Nerve Monitoring of a Non-Recurrent Laryngeal Nerve: Real-Life Data of a High-Volume Thyroid Surgery Center

非复发性喉返神经的术中连续神经监测:高容量甲状腺手术中心的真实数据

阅读:3

Abstract

Thyroid surgery is associated with a risk of injury to the recurrent laryngeal nerve, especially in the presence of anatomical variants such as a non-recurrent laryngeal nerve (NRLN). Injury to the nerve leads to transient or permanent vocal cord palsy (VCP). A novel method to prevent VCP is continuous intraoperative nerve monitoring (cIONM), but less is known about the applicability of this method in patients with NRLN. The aim of this study was to evaluate our own data regarding feasibility and detailed characteristics of cIONM in NRLN patients. We performed a monocentric retrospective cohort analysis including clinical data and intraoperative nerve monitoring data (measured by Inomed Medizintechnik GmbH, Emmendingen, 'C2' and 'C2 Xplore' device) of all thyroid surgery patients, showing NRLN between 2014 and 2022. Of 1406 patients who underwent thyroid surgery with cIONM between 2014 and 2022, 12 patients (0.9%) showed NRLN intraoperatively. Notably, cIONM was feasible in eight patients (67%). In all cases the onset latency of the right vagus nerve was shorter (<3.0 ms) than usually expected, suggesting that a short latency might be suitable to distinguish NRLN. None of the patients had a post-operative VCP. Overall, cIONM appears to be feasible and safe in NRLN patients and provides helpful information to prevent VCP.