Haemolysis induced by mechanical circulatory support devices: unsolved problems

机械循环辅助装置引起的溶血:尚未解决的问题

阅读:1

Abstract

Since the use of continuous flow blood pumps as ventricular assist devices is standard, the problems with haemolysis have increased. It is mainly induced by shear stress affecting the erythrocyte membrane. There are many investigations about haemolysis in laminar and turbulent blood flow. The results defined as threshold levels for the damage of erythrocytes depend on the exposure time of the shear stress, but they are very different, depending on the used experimental methods or the calculation strategy. Here, the results are resumed and shown in curves. Different models for the calculation of the strengths of erythrocytes are discussed. There are few results reported about tests of haemolysis in blood pumps, but some theoretical approaches for the design of continuous flow blood pumps according to low haemolysis have been investigated within the last years.