Sensitive simultaneous quantitation of testosterone and estradiol in serum by LC-MS/MS without derivatization and comparison with the CDC HoSt program

通过 LC-MS/MS 灵敏地同时定量血清中的睾酮和雌二醇,无需衍生化并与 CDC HoSt 程序进行比较

阅读:6
作者:Ryan C Schofield, Damodara R Mendu, Lakshmi V Ramanathan, Melissa S Pessin, Dean C Carlow

Background

Very sensitive measurements of serum estrogens and testosterone are important in adult and pediatric endocrinology and immunoassays are known to lack the required performance at very low levels. Our

Conclusion

We have developed an accurate and highly sensitive assay to measure E2 and Te levels in serum by HPLC-MS/MS without chemical derivatization and using commercially available calibrators.

Methods

Serum samples were prepared by the addition of internal standards followed by extraction using hexane:ethyl acetate. Chromatographic separation was achieved using a C18 column and mass spectrometry was performed in both positive and negative ion modes.

Results

The lower limits of quantitation (LLOQs) of E2 and Te were 5pg/mL and 1ng/dL, respectively. The analytical measurement range (AMR) for E2 was 5-600pg/mL and 1-1,170ng/dL for Te. Assay accuracy was determined both by comparison with a LC-MS/MS method performed at a national laboratory and the CDC HoSt program. Comparison with samples analyzed by both methods showed excellent correlation. Within-day (N=10) and between-day (N=20) CVs at concentrations spanning the AMR were less than 7% for both analytes.