Development of a permanent vacuum hollow prism air refractometer for use in dimensional metrology

研制用于尺寸计量学的永久真空空心棱镜空气折射仪

阅读:1

Abstract

Refractive index measurements are required when light is used as the basis of a measurement system. In dimensional metrology, refractive index measurements are used to compensate for the change in the speed of light. This is crucial because the SI unit for the metre is defined as the speed of light in a vacuum. Air refractometers are the most accurate way to measure the speed of light in air. Many research works to date have been performed to measure the refractive index of air using refractometers. This research uses a commercial prism as the vacuum etalon instead of the tube that is used most often. This novelty and newness of our research were to focus on the design, fabrication and testing of a refractometer which uses a permanent vacuum for ease of use but that will still have the same accuracy of other refractometers currently in use. Modifications to existing designs improved the long-term stability compared to other prism refractometers and are also potentially more accurate than tube refractometers. The results achieved with this permanent vacuum refractometer are accurate to 8.4 × 10(-8), which compares favourably with other refractometers on accuracy. It also has the added advantage that it does not require a vacuum pump, and with added laser path improved long term stability but still portable and robust enough to use in everyday applications.