Fiber-optic catheter-based polarization-sensitive OCT for radio-frequency ablation monitoring

基于光纤导管的偏振敏感光学相干断层扫描(OCT)用于射频消融监测

阅读:1

Abstract

An all-fiber optic catheter-based polarization-sensitive optical coherence tomography system is demonstrated. A novel multiplexing method was used to illuminate the sample, splitting the light from a 58.5 kHz Fourier-domain mode-locked laser such that two different polarization states, alternated in time, are generated by two semiconductor optical amplifiers. A 2.3 mm forward-view cone-scanning catheter probe was designed, fabricated, and used to acquire sample scattering intensity and phase retardation images. The system was first verified with a quarter-wave plate and then by obtaining intensity and phase retardation images of high-birefringence plastic, human skin in vivo, and untreated and thermally ablated porcine myocardium ex vivo. The system can potentially in vivo image of the cardiac wall to aid radio-frequency ablation therapy for cardiac arrhythmias.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。