Conclusions
A fast, simple, cost-effective, accurate, robust, and validated tandem mass spectrometric method was developed, and the method was successfully used for the quantitation of kynurenine pathway metabolite concentrations in hemodialysis patients.
Methods
A tandem mass spectrometric method, including protein precipitation and evaporation steps, was developed to measure serum levels of tryptophan, kynurenine, kynurenic acid, 3-hydroxykynurenine, and 3-hydroxyanthranilic acid. Samples were separated using a Phenomenex Luna C18 reversed-phase column. The kynurenine pathway metabolites were detected by tandem mass spectrometry. The developed method was validated according to Clinical & Laboratory Standards Institute (CLSI) guidelines and applied to hemodialysis samples.
Results
The developed method was linear at the concentrations of 48.8 - 25,000, 0.98 - 500, 1.2-5000, 1.2-5000, and 0.98-250 ng/mL for tryptophan, kynurenic acid, kynurenine, 3-hydroxyanthranilic acid, and 3-hydroxykynurenine, respectively. The imprecisions were less than 12 %. The median serum concentrations of tryptophan, kynurenine, kynurenic acid, 3-hydroxykynurenine, and 3-hydroxyanthranilic acid were 10530, 1100, 218, 17.6, and 25.4 ng/mL in pre-dialysis blood samples, respectively. They were 4560, 664, 135, 7.4, and 12.8 ng/mL in post-dialysis blood samples, respectively. Conclusions: A fast, simple, cost-effective, accurate, robust, and validated tandem mass spectrometric method was developed, and the method was successfully used for the quantitation of kynurenine pathway metabolite concentrations in hemodialysis patients.