[Heparin for clearance of peripherally inserted central venous catheter in newborns: an in vitro study]

[肝素用于清除新生儿外周置入中心静脉导管:一项体外研究]

阅读:1

Abstract

OBJECTIVE: To compare the efficacy of two concentrations of heparin to clear the lumen of in vitro clotted neonatal peripherally inserted central catheters (PICCs). METHODS: This is an in vitro, experimental quantitative study of 76 neonatal 2.0-Fr PICCs coagulated in vitro. The catheters were divided into two groups of 38 PICCs each. In both groups an infusion of low molecular weight heparin was administered with a dose of 25IU/mL for Group 1 and 50IU/mL for Group 2. The negative pressure technique was applied to the catheters of both groups at 5, 15 and 30min and at 4h to test their permeability. Kaplan-Meier survival analysis was used to verify the outcome of the groups according to time intervals. RESULTS: The comparison between both groups in the first 5min showed that more catheters from Group 2 were cleared compared to Group 1 (57.9 vs. 21.1%, respectively). Kaplan-Meier survival analysis showed that less time was needed to clear catheters treated with 50IU/mL of heparin (p<0.001). CONCLUSIONS: The use of low molecular weight heparin at a concentration of 50IU/mL was more effective in restoring the permeability of neonatal PICCs occluded in vitro by a clot, and the use of this concentration is within the safety margin indicated by scientific literature.