Comparison of four clinically validated testosterone LC-MS/MS assays: Harmonization is an attainable goal

四种经临床验证的睾酮 LC-MS/MS 检测方法的比较:协调是可实现的目标

阅读:6
作者:Deborah French, Julia Drees, Judith A Stone, Daniel T Holmes, J Grace van der Gugten

Conclusions

Excellent agreement between four independently developed LC-MS/MS assays demonstrates that harmonization using standard reference material is attainable. However, as we found in this study, to ensure accurate calibration it is critical to validate the concentrations of new lots of calibrators.

Methods

This study compares the

Results

Initially, one of the four assays had a mean percent difference of +11.44%, compared to the All Method Mean, but following re-verification of all five non-zero calibrator concentrations with the NIST SRM 971, the mean percent difference decreased to -4.88%. Subsequently, the agreement between all four assays showed a mean bias of <5% across the range of all testosterone concentrations (0.13-38.10 nmol/L; 3.7-1098 ng/dL), including at low concentrations of <1 nmol/L (<29 ng/dL). Conclusions: Excellent agreement between four independently developed LC-MS/MS assays demonstrates that harmonization using standard reference material is attainable. However, as we found in this study, to ensure accurate calibration it is critical to validate the concentrations of new lots of calibrators.