Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory

在哈塔伊穆斯塔法凯末尔大学临床实验室设计并验证生化检测结果自动验证系统

阅读:1

Abstract

INTRODUCTION: Autoverification (AV) is a postanalytical tool that uses algorithms to validate test results according to specified criteria. The Clinical and Laboratory Standard Institute (CLSI) document for AV of clinical laboratory test result (AUTO-10A) includes recommendations for laboratories needing guidance on implementation of AV algorithms. The aim was to design and validate the AV algorithm for biochemical tests. MATERIALS AND METHODS: Criteria were defined according to AUTO-10A. Three different approaches for algorithm were used as result limit checks, which are reference range, reference range ± total allowable error, and 2nd and 98th percentile values. To validate the algorithm, 720 cases in middleware were tested. For actual cases, 3,188,095 results and 194,520 reports in laboratory information system (LIS) were evaluated using the AV system. Cohen's kappa (κ) was calculated to determine the degree of agreement between seven independent reviewers and the AV system. RESULTS: The AV passing rate was found between 77% and 85%. The highest rates of AV were in alanine transaminase (ALT), direct bilirubin (DBIL), and magnesium (Mg), which all had AV rates exceeding 85%. The most common reason for non-validated results was the result limit check (41%). A total of 328 reports evaluated by reviewers were compared to AV system. The statistical analysis resulted in a κ value between 0.39 and 0.63 (P < 0.001) and an agreement rate between 79% and 88%. CONCLUSIONS: Our improved model can help laboratories design, build, and validate AV systems and be used as starting point for different test groups.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。