Conclusions
The measurement procedures based on UHPLC-MS/MS developed and validated in this study could be useful in pharmacokynetic and pharmacodynamic studies in subjects with liver cirrhosis who develop spontaneous bacterial peritonitis treated with ertapenem.
Methods
Samples were pretreated by acetronile protein-precipitation. Chromatographic separation is performed on a C18 reversed-phase Acquity®-UPLC®-BEHTM column (2.1 × 100 mm id, 1.7 µm) using a non-linear gradient of water/acetonitrile containing 0.1 % of formic acid at a flow rate of 0.4 mL/min. Ertapenem and its internal standard (ertapenem-D4) are detected by tandem mass spectrometry using positive electrospray ionization and multiple reaction monitoring, and using 476.2 → 346.0/432.2 as mass transition for ertapenem and 480.2 → 350.0 for its internal standard.
Results
No significant interferences or carry-over contamination were observed. Imprecisions, absolute relative bias, matrix effects and normalized recoveries were ≤14.5 %, ≤9.3 % (92.8-104.5) % and (98.8-105.8) %, respectively. Chromatographic measurement procedures were linear from (0.50-100) mg/L. Conclusions: The measurement procedures based on UHPLC-MS/MS developed and validated in this study could be useful in pharmacokynetic and pharmacodynamic studies in subjects with liver cirrhosis who develop spontaneous bacterial peritonitis treated with ertapenem.