A novel device to measure static hindlimb weight-bearing forces in pronograde rodents

一种测量啮齿动物前行性后肢静态负重力的新型装置

阅读:8
作者:Morika D Williams, Samantha L Sommer, Rachel C Meyers, Juan Valdivia, Michael W Nolan, B Duncan X Lascelles

Background

Joint pain is composed of both spontaneous and movement-induced pain. In animal models, static bodyweight distribution is a surrogate for spontaneous joint pain. However, there are no commercially-available instruments that measure static bodyweight distribution in normal, pronograde rodents. New method: We designed a Static Horizontal Incapacitance Meter (SHIM) to measure bodyweight distribution in pronograde standing rodents. We assessed the device for feasibility, repeatability, and sensitivity to quantify hindlimb bodyweight distribution. Mice and rats with unilateral inflammatory pain induced by subcutaneous injections of capsaicin or Complete Freund's Adjuvant (CFA) into the plantar surface of the left hind paw were used to measure static weight-bearing. The ability to attenuate inflammatory pain-associated weight-bearing asymmetry was tested by administering a non-steroidal anti-inflammatory drug, meloxicam.

Conclusions

The SHIM successfully detected: (a) asymmetric weight-bearing in acute and sub-chronic pain models; and (b) the analgesic effects of meloxicam. This study provides a novel tool to objectively evaluate limb use dysfunction in rodents.

Methods

In contrast with assays that measure kinetic or static weight-bearing forces (e.g., walking, or standing at a 45 ° incline), the SHIM allows evaluation of weight-bearing in rodents that are standing at rest in their normal pronograde position. Conclusions: The SHIM successfully detected: (a) asymmetric weight-bearing in acute and sub-chronic pain models; and (b) the analgesic effects of meloxicam. This study provides a novel tool to objectively evaluate limb use dysfunction in rodents.

Results

The SHIM's ability to detect significant reductions in limb loading on the injected hindlimb in mice and rats was validated using both acute and sub-chronic pain models. Treatment with meloxicam partially reversed CFA-induced effects. Comparison with existing methods: In contrast with assays that measure kinetic or static weight-bearing forces (e.g., walking, or standing at a 45 ° incline), the SHIM allows evaluation of weight-bearing in rodents that are standing at rest in their normal pronograde position. Conclusions: The SHIM successfully detected: (a) asymmetric weight-bearing in acute and sub-chronic pain models; and (b) the analgesic effects of meloxicam. This study provides a novel tool to objectively evaluate limb use dysfunction in rodents.