SERIAL transmit - parallel receive (ST(x)PR(x)) MR imaging produces acceptable proton image uniformity without compromising field of view or SAR guidelines for human neuroimaging at 9.4 Tesla

串行发射-并行接收 (ST(x)PR(x)) 磁共振成像技术可在 9.4 特斯拉下产生可接受的质子图像均匀性,且不会影响视野或 SAR 值,符合人体神经成像的指导原则。

阅读:2

Abstract

PURPOSE: Non-uniform B1(+) excitation and high specific absorption rates (SAR) compromise proton MR imaging of human brain at 9.4 T (400.5 MHz). By combining a transmit/receive surface coil array using serial transmission of individual coils with a total generalized variation reconstruction of images from all coils, acceptable quality human brain imaging is demonstrated. METHODS: B(0) is shimmed using sodium MR imaging (105.4 MHz) with a birdcage coil. Proton MR imaging is performed with an excitation/receive array of surface coils. The modified FLASH pulse sequence transmits serially across each coil within the array thereby distributing SAR in time and space. All coils operate in receive mode. Although the excitation profile of each transmit coil is non-uniform, the sensitivity profile estimated from the non-transmit receive coils provides an acceptable sensitivity correction. Signals from all coils are combined in a total generalized variation (TGV) reconstruction to provide a full field of view image at maximum signal to noise (SNR) performance. RESULTS: High-resolution images across the human head are demonstrated with acceptable uniformity and SNR. CONCLUSION: Proton MR imaging of the human brain is possible with acceptable uniformity at low SAR at 9.4 Tesla using this serial excitation and parallel reception strategy with TGV reconstruction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。