Quantitative RNA assessment and long-term stability in the FFPE tumor samples using Digital Spatial Profiler

使用数字空间分析器对 FFPE 肿瘤样本中的 RNA 进行定量评估和长期稳定性

阅读:5
作者:S Gupta, T Chen, B Destenaves

Background

Long-term storage of tissue slides has been reported to induce reduced biomarker (e.g. proteins and messenger RNA) detection. This study aimed to evaluate the impact of long-term storage time (0, 16, 24 and 36 weeks) and treatment conditions (non-paraffin and paraffin dipping) at 4°C on RNA quality in formalin-fixed and paraffin-embedded (FFPE) tissue sections. Materials and

Conclusion

Although both treatment conditions provided similar results in terms of count reproducibility and signal preservation, we recommend paraffin dipping to generate reproducible RNA results and optimize sample storage. Technology behind the NanoString GeoMx DSP platform shows a robust and reproducible RNA signal from multiple targets in the FFPE tissue sections stored at 4°C for at least up to 36 weeks.

Methods

NanoString GeoMx Digital Spatial Profiling (DSP), a novel platform that allows spatial profiling, was used to profile RNA in human bladder cancer FFPE tissue sections.

Results

We observed excellent consistency of quantitative DSP RNA counts of all targets between two different treatment conditions (R > 0.97, Pearson correlation) at each time point and among all four different storage time points (R > 0.96, Pearson correlation) within each treatment condition. No significant difference was observed in the percentage of target genes with sufficient signal across two different treatment conditions at any time point (0 week, P = 0.96; 16 weeks, P = 0.76; 24 weeks, P = 0.96; 36 weeks, P = 0.76, Kolmogorov-Smirnov test) and across all four different storage time points (P > 0.05, Kolmogorov-Smirnov test) in either treatment condition.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。