Measurement of the microscopic viscosities of microfluids with a dynamic optical tweezers system

利用动态光镊系统测量微流体的微观粘度

阅读:1

Abstract

Viscosity coefficients of microfluids-Newtonian and non-Newtonian-were explored through the rotational motion of a particle trapped by optical tweezers in a microflute. Unlike conventional methods based on viscometers, our microfluidic system employs samples of less than 30 µl to complete a measurement. Viscosity coefficients of ethanol and fetal bovine serum, as typical examples of Newtonian and non-Newtonian fluids, were obtained experimentally, and found to be in excellent agreement with theoretical predictions. Additionally, a practical application to a DNA solution with incremental ethidium bromide content was employed and the results are consistent with clinical data, indicating that our system provides a potentially important complementary tool for use in such biological and medical applications.