The interplay of electrode- and bio-materials in a redox-cycling-based clozapine sensor

基于氧化还原循环的氯氮平传感器中电极材料和生物材料的相互作用

阅读:1

Abstract

We investigate gold, TiN, and platinum in combination with a chitosan-catechol-based redox-cycling system (RCS) for electrochemical detection of the antipsychotic clozapine. We have previously demonstrated the RCS for detection of clozapine in serum, but challenges remain regarding low signal-to-noise ratios. This can be mitigated by selection of electrode materials with beneficial surface morphologies and/or compositions. We employ cyclic voltammetry to assess the redox current generated by clozapine, and differentiate solely surface-area-based effects from clozapine-specific ones using a standard redox couple. We find that nano- and microstructured platinum greatly amplifies the clozapine signal compared to gold (up to 1490-fold for platinum black). However, the material performs poorly in the presence of chloride ions, and RCS modification provides no further amplification. The RCS combined with atomic-layer-deposited (ALD) TiN, on the other hand, increases the signal by 7.54 times, versus 2.86 times for RCS on gold, with a 9.2-fold lower variability, indicating that the homogenous and chemically inert properties of ALD-TiN may make it an ideal electrode material.