Mechanical strength of a new plate compared to six previously tested opening wedge high tibial osteotomy implants

与之前测试过的六种开放楔形高位胫骨截骨植入物相比,新型钢板的机械强度

阅读:1

Abstract

BACKGROUND: This study aimed to assess the mechanical static and fatigue strength provided by the FlexitSystem plate in medial opening wedge high tibial osteotomies (MOWHTO), and to compare it to six previously tested implants: the TomoFix small stature, the TomoFix standard, the ContourLock, the iBalance, the second generation PEEKPower and the size 2 Activmotion. Thus, this will provide surgeons with data that will help in the choice of the most appropriate implant for MOWHTO. METHODS: Six fourth-generation tibial bone composites underwent a MOWHTO and each was fixed using six FlexitSystem plates, according to standard techniques. The same testing procedure that has already been previously defined, used and published, was used to investigate the static and dynamic strength of the prepared bone-implant constructs. The test consisted of static loading and cyclical loading for fatigue testing. RESULTS: During static testing, the group constituted by the FlexitSystem showed a fracture load higher than the physiological loading of slow walking (3.7 kN > 2.4 kN). Although this fracture load was relatively small compared to the average values for the other Implants from our previous studies, except for the TomoFix small stature and the Contour Lock. During fatigue testing, FlexitSystem group showed the smallest stiffness and higher lifespan than the TomoFix and the PEEKPower groups. CONCLUSIONS: The FlexitSystem plate showed sufficient strength for static loading, and average fatigue strength compared to the previously tested implants. Full body dynamic loading of the tibia after MOWHTO with the investigated implants should be avoided for at least 3 weeks. Implants with a wider T-shaped proximal end, positioned onto the antero-medial side of the tibia head, or inserted in the osteotomy opening in a closed-wedge construction, provided higher mechanical strength than implants with small a T-shaped proximal end, centred onto the medial side of the tibia head.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。