Assessment of the Diagnostic Performance of Fully Automated Hepatitis E Virus (HEV) Antibody Tests

全自动戊型肝炎病毒(HEV)抗体检测诊断性能评估

阅读:1

Abstract

The detection of anti-hepatitis E virus (HEV) antibodies contributes to the diagnosis of hepatitis E. The diagnostic suitability of two automated chemiluminescence immunoassays (CLIAs, LIAISON(®) MUREX Anti-HEV IgG/Anti-HEV IgM test, DiaSorin) was assessed by comparison with the results of a combination of enzyme immunoassays and immunoblots (recomWell HEV IgG/IgM ELISA, recomLine HEV IgG/IgM, MIKROGEN). Samples with a deviating result were analyzed with the WANTAI ELISAs. Compared to the recomWell ELISAs, the Anti-HEV IgG CLIA had a percentage overall agreement (POA) of 100% (149/149; 95% CI: 97.5-100%) and the Anti-HEV IgM CLIA had a POA of 83.3% (85/102; 95% CI: 74.9-89.3%); considering the recomLine HEV IgM results, the POA was 71.7% (38/53; 95% CI: 58.4-82%). The WANTAI test confirmed 52.9% (9/17) of negative CLIA IgMs; HEV RNA was not detectable. Since acute infection with the Epstein-Barr virus (EBV) or human cytomegalovirus (CMV) may influence the results of other serological assays, HEV antibodies were examined in 17 EBV and 2 CMV patients: One had an isolated and probably unspecific HEV IgM in the CLIA, as HEV RNA was not detectable. Both CLIAs are well suited for HEV diagnostics, but isolated IgM should be confirmed. An acute EBV/CMV infection can influence HEV serodiagnostics.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。