Tracking the Traveled Distance of Capsule Endoscopes along a Gastrointestinal-Tract Model Using Differential Static Magnetic Localization

利用差分静态磁定位技术追踪胶囊内镜在胃肠道模型中的行进距离

阅读:2

Abstract

The traveled distance and orientation of capsule endoscopes for each video frame are not available in commercial systems, but they would be highly relevant for physicians. Furthermore, scientific approaches lack precisely tracking the capsules along curved trajectories within the typical gastrointestinal tract. Recently, we showed that the differential static magnetic localisation method is suitable for the precise absolute localisation of permanent magnets assumed to be integrated into capsule endoscopes. Thus, in the present study, the differential method was employed to track permanent magnets in terms of traveled distance and orientation along a length trajectory of 487.5 mm, representing a model of the winding gastrointestinal tract. Permanent magnets with a diameter of 10 mm and different lengths were used to find a lower boundary for magnet size. Results reveal that the mean relative distance and orientation errors did not exceed 4.3 ± 3.3%, and 2 ± 0.6∘, respectively, when the magnet length was at least 5 mm. Thus, a 5 mm long magnet would be a good compromise between achievable tracking accuracy and magnet volume, which are essential for integration into small commercial capsules. Overall, the proposed tracking accuracy was better than that of the state of the art within a region covering the typical gastrointestinal-tract size.