Disconnection of a stent retriever's pushwire caught by an accordion-like deformed aspiration catheter during mechanical thrombectomy: illustrative case

机械取栓术中支架取栓器推线被褶皱状变形的抽吸导管卡住而脱落:病例分析

阅读:2

Abstract

BACKGROUND: In mechanical thrombectomy for acute large vessel occlusion, a combined technique of using both a stent retriever and an aspiration catheter has been widely used. The authors report a case in which a stent retriever's pushwire and a microcatheter were caught and disconnected by an accordion-like deformed aspiration catheter. OBSERVATIONS: A 74-year-old man underwent mechanical thrombectomy for a left M1 occlusion. A stent retriever was deployed from the left M2 to the left distal M1, and an aspiration catheter was advanced to the left distal M1. When the stent retriever and microcatheter were pulled into the aspiration catheter at the distal M1 without releasing the deflection, traction resistance of the stent retriever occurred, and the aspiration catheter contracted and deformed like an accordion distal to the tip of the guiding catheter. The stent retriever's pushwire and the microcatheter were caught and disconnected. LESSONS: When a stent retriever is pulled into a flexible aspiration catheter in a case with vascular tortuosity, it may be caught by an accordion-like deformed aspiration catheter and disconnected. It is necessary to release the deflection of the aspiration catheter once traction resistance of the stent retriever and deflection of the aspiration catheter occur.