Long-term reliability of sweet-tip type screw-in leads

甜尖型螺纹导线的长期可靠性

阅读:2

Abstract

BACKGROUND: Active fixation leads have provided stable atrial and ventricular pacing; however, long-term follow-up data have not been satisfactory. The purpose of this study was to investigate the long-term reliability of active fixation leads and their electrical characteristic stability. METHODS: A total of 1196 pacing leads were implanted in 830 patients consecutively between 2002 and 2013. In this retrospective study, we were able to trace 1092 leads in 750 patients to investigate the prognosis of implanted leads. The measurement values (including pacing thresholds, sensing amplitudes, and lead impedances of both the atrial and ventricular leads) were obtained from medical records at the time of implantation and during follow up at the outpatient device clinic. All pacing leads were FINELINE II Sterox EZ Leads (Boston Scientific, MN, USA), which are sweet-tip type screw-in active fixation leads, except for the shock leads in patients with implantable cardioverter defibrillator. RESULTS: The mean follow-up period was 51.3±29.2 months (median, 48 months). A total of 1092 leads were implanted in either the atrium (682 leads) or the ventricle (410 leads). Venous access was achieved through cephalic vein cut down (CVC) method (914 leads) or the subclavian vein puncture (SVP) method (178 leads). The overall lead survival rate was 99.6% at both 5 and 10 years. Lead fracture was observed in 4 of 1092 leads (0.37%), all of which were implanted by the SVP method. No lead fracture occurred among patients wherein CVC method was applied (p<0.01). Device-related infection was observed in four patients (0.53%). CONCLUSIONS: The overall reliability and stability of sweet-tip type screw-in leads were satisfactory throughout the long-term follow-up period (median, 4 years). Because it was associated with less lead fractures, cut-down access from the cephalic vein may be recommended as the first-line approach when considering the importance of long-term durability of pacing leads.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。