Ophthalmologists' assessment of the handling characteristics of the novel Finesse Reflex Handle in comparison to those of a conventional handle

眼科医生对新型Finesse Reflex手柄与传统手柄的操作特性进行了比较评估

阅读:1

Abstract

Internal limiting membrane (ILM) peeling requires a delicate handling technique. It is also important that ophthalmologists can use the ILM forceps handle of their preference. This study objectively and subjectively evaluated the handling of the novel Finesse Reflex Handle (Reflex) in comparison with that of a conventional handle. The force required to close the forceps tips, evaluated using a digital force gauge, was significantly lesser for Reflex than for the conventional handle (3.14 ± 0.09 N vs. 3.84 ± 0.06 N, P < 0.001). Twenty-one ophthalmologists with various levels of experience answered a questionnaire after using both handles, and the total questionnaire score for Reflex was higher than that for the conventional handle (35.0 ± 3.7 vs. 30.0 ± 6.9, P = 0.01). Furthermore, the duration of experience as an ophthalmologist was negatively correlated with the vertical motion, assessed by video analysis, for the conventional handle (P = 0.02, r =  - 0.50) but not for Reflex (P = 0.26). In conclusion, objective and subjective analyses revealed that compared with the conventional handle, the novel Reflex handle had more favourable handling characteristics. Most ophthalmologists preferred the handling of Reflex. Reflex may compensate for a lack of surgical experience.