Clinical Utility of Artificial Intelligence Algorithms to Enhance Wide-Field Optical Coherence Tomography Angiography Images

人工智能算法在增强宽视野光学相干断层扫描血管造影图像方面的临床应用

阅读:1

Abstract

The aim of this paper is to investigate the clinical utility of the application of deep learning denoise algorithms on standard wide-field Optical Coherence Tomography Angiography (OCT-A) images. This was a retrospective case-series assessing forty-nine 10 × 10 mm OCT-A1 macula scans of 49 consecutive patients attending a medical retina clinic over a 6-month period. Thirty-seven patients had pathology; 13 had none. Retinal vascular layers were categorised into superficial or deep capillary plexus. For each category, the retinal experts compared the original standard image with the same image that had intelligent denoise applied. When analysing the Superficial Capillary Plexus (SCP), the denoised image was selected as "best for clinical assessment" in 98% of comparisons. No difference was established in the remaining 2%. On evaluating the Deep Capillary Plexus (DCP), the denoised image was preferred in 35% of comparisons. No difference was found in 65%. There was no evidence of new artefactual features nor loss of anatomical detail in denoised compared to the standard images. The wide-field denoise feature of the Canon Xephilio OCT-A1 produced scans that were clinically preferable over their original OCT-A images, especially for SCP assessment, without evidence for causing a new artefactual error.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。