Determination of the appropriate physical density of internal metallic ports in temporary tissue expanders for the treatment planning of post-mastectomy radiation therapy

确定临时组织扩张器内部金属端口的合适物理密度,以用于乳房切除术后放射治疗的治疗计划

阅读:1

Abstract

Some patients undergoing breast reconstruction require post-mastectomy radiation therapy, but the metallic ports used in temporary tissue expanders attenuate the X-rays. In this study, we evaluated by the film method, the attenuation of 4 MV and 6 MV X-rays after passing through a metallic port, with the aim of identifying a useful method for determining the appropriate density to use in the radiation treatment planning system (RTPS), taking into account the distance between the metallic port and the targets. Radiochromic film was used to measure depth doses after the X-rays passed through the metallic port. The physical density allotted to the metal port portion was varied on the RTPS within the range 1-16 g/cm3, and the physical density values were calculated that best reproduced the depth-dose distribution extrapolated from the film method. When the metallic port was orientated perpendicularly, the attenuation of the X-rays peaked at ~7% at both 4 MV and 6 MV. In the parallel orientation, the X-rays were attenuated by up to ~40% at 4 MV and by up to ~30% at 6 MV. We estimated the optimum physical density to be 9.8 g/cm3, which yielded the best fit with the actual measurements. We demonstrated the most likely range for the target depth from the CT images of actual patients and, within this range, we identified the optimum physical density at which the measured and calculated values were most consistent with each other.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。